Noninvasive Assessment of Arterial Pulse-Pressure Variation During General Anesthesia: Clinical Evaluation of a New High-Fidelity Upper Arm Cuff

  • Philippe Conter
  • , Josef Briegel
  • , Torsten Baehner
  • , Alois Kreitmeier
  • , Agnes S Meidert
  • , Martin Tholl
  • , Franz Schwimmbeck
  • , Andreas Bauer
  • , Ulrich J Pfeiffer
  • Masaryk University and University Hospital Brno
  • Department of Anesthesiology
  • Philips Business Unit Hospital Patient Monitoring

Research output: Contribution to journalOriginal Articlepeer-review

Abstract

OBJECTIVES: To compare noninvasive pulse-pressure variation (PPV) measurements obtained from a new high-fidelity upper arm cuff using a hydraulic coupling technique to corresponding intraarterial PPV measurements.

DESIGN: The authors used prospective multicenter comparison and development studies for the new high-fidelity upper arm cuff.

SETTING: The study was performed in the departments of Anesthesiology at the Ludwig-Maximilians-Universität München Hospital, the University Hospital of Bonn, and the RoMed Hospital in Rosenheim (all Germany).

PARTICIPANTS: A total of 153 patients were enrolled, undergoing major abdominal surgery or neurosurgery with mechanical ventilation. For the evaluation of PPV, 1,467 paired measurements in 107 patients were available after exclusion due to predefined quality criteria.

INTERVENTIONS: Simultaneous measurements of PPV were performed from a reference femoral arterial catheter (PPVref) and the high-fidelity upper arm cuff (PPVcuff). The new device uses a semirigid conical shell. It incorporates a hydraulic sensor pad with a pressure transducer, leading to a tissue pressure-pulse contour with all characteristics of an arterial- pulse contour.

MEASUREMENTS AND MAIN RESULTS: The comparative analysis of the included measurements showed that PPVref and PPVcuff were closely correlated (r = 0.92). The mean of the differences between PPVref and PPVcuff was 0.1 ± 2.0%, with 95% limits of agreement between -4.1% and 3.9%. To track absolute changes in PPV >2%, the concordance rate between the 2 methods was 93%.

CONCLUSIONS: The new high-fidelity upper arm cuff method provided a clinically reliable estimate of PPV.

Original languageEnglish
Pages (from-to)1382-1389
Number of pages8
JournalJOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA
Volume37
Issue number8
DOIs
Publication statusPublished - Aug 2023

Keywords

  • Humans
  • Arm
  • Blood Pressure Determination/methods
  • Prospective Studies
  • Blood Pressure
  • Anesthesia, General