Abstract
This study investigates the crucial factors influencing the end-systolic and end-diastolic volumes in MRI volumetry and their direct effects on the derived functional parameters. Through the simultaneous acquisition of 2D-cine and 3D whole-heart slices in end-diastole and end-systole, we present a novel direct comparison of the volumetric measurements from both methods. A prospective study was conducted with 18 healthy participants. Both 2D-cine and 3D whole-heart sequences were obtained. Despite the differences in the creation of 3D volumes and trigger points, the impact on the LV volume was minimal (134.9 mL +/- 16.9 mL vs. 136.6 mL +/- 16.6 mL, p < 0.01 for end-diastole; 50.6 mL +/- 11.0 mL vs. 51.6 mL +/- 11.2 mL, p = 0.03 for end-systole). In our healthy patient cohort, a systematic underestimation of the end-systolic volume resulted in a significant overestimation of the SV (5.6 mL +/- 2.6 mL, p < 0.01). The functional calculations from the 3D whole-heart method proved to be highly accurate and correlated well with function measurements from the phase-contrast sequences. Our study is the first to demonstrate the superiority of 3D whole-heart volumetry over 2D-cine volumetry and sheds light on the systematic error inherent in 2D-cine measurements.